Abstrato

Gas chromatographic-mass spectrometric determination of tetracycline in biological fluids

S.Thangadurai


A gas chromatographic-mass spectrometric (GC-MS) method is described for the determination of tetracycline (TC) residue in biological fluids. This method allows detection of residual TC in biological fluids by using singleionmonitoring (SIM), confirmation by a full scan electron impact (EI) mass spectrum is possible if residual level in a sample is >1 µg/mL. TC is extracted with chloroformfroma sample and cleaned up by n-hexane washing followed by partition between chloroform and phosphate buffer solution. The cleaned up extract is acetylated in acetic anhydride-pyridinemixture (1: 2) at room temperature. The reaction mixture is injected into the GC-MS apparatus, and the detection has been conducted using SIM at m/z 149. The detection limit is 0.01µg/mL. This method could be adopted in the forensic laboratories as well as in the toxicological laboratories.


Indexado em

  • CASS
  • Google Scholar
  • Abra o portão J
  • Infraestrutura Nacional de Conhecimento da China (CNKI)
  • CiteFactor
  • Cosmos SE
  • Diretório de indexação de periódicos de pesquisa (DRJI)
  • Laboratórios secretos de mecanismos de pesquisa
  • Fator de impacto do artigo acadêmico (SAJI))
  • ICMJE

Veja mais

Flyer