Abstrato

Ion-selective electrode for zinc determination in pharmaceutical analysis

S.K.Mohamed


An optimal composition for liquidmembrane chelating potential phase of a zinc function liquid electrode was established. Its basic analytical parameters, such as the slope of characteristics ( 29.5 mV pa Zn -1 ), the detection limit ( 7.3 x 10-6mol L-1), lifetime ( 50 days ), response time ( 12 s ), selectivity against some metal ions, as well as the dependence of the electrode potential on pH, were established. The electrode was used to determine the presence of zinc ions in pharmaceutical formulations. The results obtained were compared with the determination made by use of a voltammetricmethod.


Isenção de responsabilidade: Este resumo foi traduzido usando ferramentas de inteligência artificial e ainda não foi revisado ou verificado

Indexado em

  • CASS
  • Google Scholar
  • Abra o portão J
  • Infraestrutura Nacional de Conhecimento da China (CNKI)
  • CiteFactor
  • Cosmos SE
  • Biblioteca de Periódicos Eletrônicos
  • Diretório de indexação de periódicos de pesquisa (DRJI)
  • Laboratórios secretos de mecanismos de pesquisa
  • ICMJE

Veja mais

Flyer