Abstrato

Reverse Phase HPLC Method For the Analysis of Alfuzosin Hydrochloride in Pharmaceutical Dosage Forms

S. Appala Raju, Arvind B. Karadi and Shobha Manjunath


A simple, rapid and reproducible high performance reversed phase liquid chromatographic method has been developed for the estimation of alfuzosin hydrochloride in bulk drug samples and pharmaceutical dosage forms using RP C-18 column. The mobile phase consists of buffer (pH 3.8) and acetonitrile in the ratio 650 : 350 (v/v), respectively, and was pumped at 1.0 mL/min at 30°C. The detection was carried out at 244 nm and the calibration curve was linear in the range of 0.02 to 20 μg/mL. The method was statistically validated for its linearity, precision and accuracy. The intra- and inter-day variation was found to be less than 1% showing high precision of the assay method. Parameters of validation obtained prove the accuracy of the method and its applicability for the determination of alfuzosin hydrochloride in tablet formulations.


Isenção de responsabilidade: Este resumo foi traduzido usando ferramentas de inteligência artificial e ainda não foi revisado ou verificado

Indexado em

  • CASS
  • Google Scholar
  • Abra o portão J
  • Infraestrutura Nacional de Conhecimento da China (CNKI)
  • CiteFactor
  • Cosmos SE
  • Diretório de indexação de periódicos de pesquisa (DRJI)
  • Laboratórios secretos de mecanismos de pesquisa
  • ICMJE

Veja mais

Flyer