Abstrato

RP-HPLC Estimation of Tadalafil in Tablet Dosage Form

A. S. Sutar, C. S. Magdum, S. S. Patil and N. S. Naikawadi


A simple, specific, accurate and precise method, reverse phase high performance liquid chromatography was developed for estimation of tadalafil in tablet dosage form. For HPLC method, Hi- Qsil C18-10 column in isocratic mode with mobile phase containing acetonirtile; acetate buffer pH (2.8) in ratio of 45 : 55 v/v was used. The flow rate was 1 mL/min and effluent was monitored at 283 nm. Retention time was found to be 4.46 ± 0.03 minutes. Method was validated in terms of linearity, accuracy and precision. Results of the analysis show that the method can be used for the estimation of tadalafil in tablet dosage form.


Isenção de responsabilidade: Este resumo foi traduzido usando ferramentas de inteligência artificial e ainda não foi revisado ou verificado

Indexado em

  • CASS
  • Google Scholar
  • Abra o portão J
  • Infraestrutura Nacional de Conhecimento da China (CNKI)
  • CiteFactor
  • Cosmos SE
  • MIAR
  • Laboratórios secretos de mecanismos de pesquisa
  • Euro Pub
  • Universidade de Barcelona
  • ICMJE

Veja mais

Flyer